98 Chapter 2
S:\agilent\83217\OPT001\BOOK\CHAPTERS\tdscrip.fm
Test Descriptions
TEST_46 - CPD Talk Time
TEST_46 - CPD Talk Time
Description
This test measures the mobile station battery life/talk time by
simulating the battery load conditions experienced in a typical system.
1. Before a call is started, two questions are asked:
“Test urban topography or suburban topography?”
& “Loopback data mode or echo data mode?”
Where both questions are answered by pressing the appropriated
USER key K1 or K2. The first question determines which mobile
station transmit power statistics will be used during the test as
defined by CDG Stage 4 Optional System performance Tests
Revision 2.0. The second question determines which service option
and data type or data rate in the CDMA CALL CONTROL screen
will be used. Selecting loopback data mode sets Service Option 2 if
parameter 19. RT Rate Set Selection [0=9600 1=14400] is set to 0
(rate set = 9600) or Service Option 9 if parameter 19. RT Rate Set
Selection [0=9600 1=14400] is set to 1(rate set =14400). It also sets
the data rate to ‘random’ which statistically spends an equal amount
of time at each of the four data rates (about 47% on 53% off).
Selecting echo data mode sets Service Option 1 if parameter 19. RT
Rate Set Selection [0=9600 1=14400] is set to 0 (rate set = 9600) or
Service Option 32768 if parameter 19. RT Rate Set Selection
[0=9600 1=14400] is set to 1 (rate set = 14400)). It also sets the data
type to ‘echo’ where parameter 10. CPD Echo Delay [0, 2 or 5]
(Seconds) sets the echo delay. This allows the user to couple custom
voice data into the mobile for voice echo.
NOTE
Parameter 10. CPD Echo Delay [0, 2 or 5] (Seconds) is used if you have
a Agilent 8924C or Agilent 8924E with firmware A.02.03 or later.
2. The correct type of service option call as defined in step 1 above is
attempted, unless that type of call is already connected. If the MSUT
has not registered, the call attempt may be unsuccessful and the
error message “Failed to establish a call in test <>” will be displayed.
To correct this problem, perform the CPD Registration test.
3. The forward link power is then cycled through a statistical profile to
simulate actual system use.The Sector A power level is displayed at
each new power level.
4. This is continued until the mobile station ends the call and the test
software detects an GPIB interrupt to determine when the connected
state is false.